SEM METHODS
PROTOCOLS
FOR SEM TISSUE PREPARATION
The
methods outlined below for tissue fixation are very general and should not be
used in place of methods published for the tissue you are examining. For people
experienced in TEM tissue preparation, the initial stages of fixation and
dehydration to absolute ethanol are identical except that larger pieces of
tissue may be used.
CAUTION:
Glutaraldehyde and osmium tetroxide are possible carcinogens. Gloves must be
worn when handling these chemicals. The chemicals must be used in designated
areas of the hood in SCST147 and disposed of in the proper containers. Pasteur
pipets that contact these chemicals should also be disposed of in the labeled
container in the hood.
Fixation:
1.
Tissue should be immersed in 3% buffered glutaraldehyde for 2-3 hours (fixation
times of up to 24 hours may be necessary for some tissues). Make sure that the
appropriate label accompanies each vial of tissue you prepare.
2.
Wash 2-3 times in buffer (10-15 mins each change) to remove all traces of
glutaraldehyde.
3.
Post-fix in 2% osmium tetroxide in buffer for 1-2 hours for animal tissue and
overnight (12-18 hours) for plant tissue. Post fixation may be omitted in some
protocols for SEM.
4.
Dehydrate tissue through graded ethanols:
50%..............15-30 mins
70%..............15-30 mins (tissue
may be left here)
80%..............15-30 mins
95%..............30 mins
100%.............30 mins
100%.............30 mins
5.
The tissue may now be transferred to the critical point dryer, but some workers
recommend the use of a transitional fluid such as amyl acetate.
6.
For use of amyl acetate, make dilutions of amyl acetate and 100% ethanol. You
will need 30% amyl acetate, 60%, and
90%. Process the tissue for 10 mins each through each dilution, and then
through 2 changes (10 mins each) of 100% amyl acetate.
USE OF THE CPD2 CRITICAL POINT DRYER
CAUTION: Do not use
this instrument if you are unsure of any of the steps outlined below. This instrument operates under high
pressure which is dangerous for the uninformed user.
1.
Your tissue should be immersed in alcohol or the transitional fluid. Transfer
the specimen to a specimen basket. Put the basket in alcohol or transitional
fluid to keep it moist.
2.
On the CRITCAL POINT DRYER: Check to see that the VENT, DRAIN and FILL valves
are closed. DO NOT OVERTIGHTEN THESE.
3.
Depress the POWER SWITCH and the VALVE HEATER. The buttons should light.
4.
Depress CHAMBER ILLUMINATION.
5.
Open the main valve on the CO2
tank slightly to fill the charging hose, then open the FILL valve and
watch as liquid CO2 enters
the chamber.
6.
Open the VENT valve slightly. By venting the CO2 you will begin to cool the chamber.
7.
Observe the temperature gauge. You will
need to experiment with opening the FILL valve and VENT valve alternately
filling the chamber 1/2 to 2/3 and then venting to cool the chamber. The
pressure readout should be about 800 psi. Use this procedure to cool the
chamber to 10-15oC.
8.
Remove the specimen basket from the alcohol or transitional fluid. Place the
basket in the carrier and install the carrier in the chamber. Do not be alarmed
if the chamber warms slightly during this procedure. NOTE: If you overtighten
the chamber cover at this point it will be extremely difficult to remove at the
end of your run.
9.
If you wish to place transitional fluid in the chamber, consult Dr. Baird or
the user's manual.
10.
Close the VENT valve and open the FILL valve to fill the chamber with enough CO2
to cover the basket(s) you used. Close the FILL valve. Allow the tissue
to "soak" in the CO2 for a few minutes.
11.
Partially open the DRAIN valve and lower the CO2 level to just above the surface of the
sample (you will have to estimate this, but the sample has to be submerged at
all times). Close the DRAIN valve.
12.
Open the FILL valve and admit more CO2 into the chamber to cover the basket. Close
the FILL valve. Repeat the soak.
13.
Repeat step 10.
14.
Open the FILL valve to fill the chamber enough to fully cover the baskets (1/2
to 2/3 full in most cases).
15.
Close all valves. Remember to close the main valve on the CO2 tank.
16.
Depress the CHAMBER HEATER switch. The temperature and pressure should begin to
rise and the CO2 will appear
to boil.
17.
The pressure will begin to climb to its set point of 1200 psi and the
temperature to its set point of 42oC. The machine will probably
reach the pressure set point first.
18.
When a pressure of 1200 psi is reached, the heater light will click off. Open
the VENT valve slightly (then close it) to lower the pressure to about 1100,
and the light will come back on. Watch the pressure and temperature balance -
the chamber is not heating when the pressure is above 1200 and the light is off!
19.
When the temperature set point of 42oC is reached, the CHAMBER
HEATER light will go off. All of the CO2 in the chamber should be gone.
20.
Open the VENT valve to begin to reduce the pressure in the chamber by about 100
psi per minute. Note that the VENT RATE valve will help you regulate this. It
is important to vent slowly to avoid condensation.
21.
When the pressure is ZERO, turn off the CHAMBER HEATER, VALVE HEATER, CHAMBER
ILLUMINATION and then the POWER.
22.
Open the VENT and DRAIN valves.
23.
Loosen the cover on the chamber and remove your samples. Note that the samples
will be fragile.
24.
Samples may now be affixed to the stubs using the appropriate mounting medium.
If in doubt about this, ASK.
25.
If you are not going to use the sputter coater, clean up and lock the lab.
WHEN
YOU ARE FINISHED, BE SURE THAT ALL PARTS ARE RETURNED TO THEIR PROPER CABINETS
AND ALL MACHINES TURNED OFF. CLEAN ANY AREAS YOU HAVE USED. FAILURE TO DO SO
WILL RESULT IN SUSPENSION OF YOUR SEM LAB PRIVILEGES.
LMB 3/99
CAUTION: Do not use
this instrument if you are unsure of any of the steps outlined below. This
instrument employs high voltage which is dangerous for the uninformed user.
1.
Open the regulator valve for the argon. NOTE: The main tank valve on the argon
tank is always open. The operating pressure on the tank should be 2-5 on the
left hand gauge.
2.
Carefully remove the metal top plate and set it aside. If desired, the glass
cylinder may also be removed - again, be careful.
3.
Place the stubs in the specimen stage.
4.
Replace the glass cylinder and the top plate.
5.
Set the timer on the front to the desired setting (if unsure, start with 40
sec).
6.
Check the GAS ADJUST valve on the side of the chamber to be sure it is closed
(clockwise). DO NOT OVERTIGHTEN VALVE, YOU WILL RUIN IT.
7.
Depress the POWER SWITCH to evacuate the chamber. You will note the pressure
drop on the vacuum gauge.
8.
Open the GAS ADJUST valve to admit argon to the chamber. Allow the pressure to
rise to 1-2 mBar and to remain there for 10 sec. Then close the gas adjust
valve and allow the vacuum to be reestablished. This is called
"flushing". Flush chamber for at least 2 cycles.
9.
Close the GAS ADJUST valve and pump the chamber to below 0.1 mBar.
10.
Open the GAS ADJUST valve to the pressure rises to between 0.1-0.2 mBar.
11.
Depress the READY/TEST switch momentarily and read the discharge on the plasma
current gauge. The current should be about 18 mA. NOTE: If the current is too
high the needle will arc, and the reset light will illuminate. Press RESET,
evacuate some of the argon, and try again. The high current is usually the
result of too much argon.
12.
When the current is about 18 mA you should see a lavender glow in the chamber.
13.
Depress the START switch to begin sputtering.
14.
At the end of the sequence, the plasma current will turn off. Depress the POWER
SWITCH to discontinue pumping.
15.
Admit air to the vacuum chamber with the vent valve on the top of the chamber.
16.
Remove your specimens and replace the top plate.
17.
Close the argon valve.
18.
Lock the lab.
WHEN
YOU ARE FINISHED, BE SURE THAT ALL PARTS ARE RETURNED TO THEIR PROPER CABINETS,
AND ALL MACHINES TURNED OFF. CLEAN ANY AREAS YOU HAVE USED. FAILURE TO DO SO WILL
RESULT IN SUSPENSION OF YOUR SEM LAB PRIVILEGES.
LMB 3/99
NOTE:
These basic instructions are intended for individuals who have previously been
trained on the use of the SEM. If you have questions at any point, please ask.
Basic
Lab Rules:
OPERATION
OF THE MICROSCOPE
Use
of Automatic Modes:
TO
INSERT A SPECIMEN:
TO VIEW YOUR
SPECIMEN:
a. Turn the kV off, the red button will
illuminate.
b. Press the STANDBY button and do
whatever you need to do.
c. When you return, press EVACUATE and
wait for the READY light. You may now turn on the kV and resume your
observations.
TO SHUT DOWN
THE MICROSCOPE:
CAUTION: These steps
must be done in the order listed - don't rush these steps!
PHOTOGRAPHY:
Using
the video printer: Make sure the printer is ON. There is a switch on the back
of the printer to turn it on and the power switch on the front must also be
depressed. When the printer is ready the green icons on the front of the
printer will be illuminated.
Using
the Polaroid Camera: Unless you are in Biology 132, you must provide your own
film. Polaroid type 52 gives only prints, Polaroid type 55 also gives a
negative if you process it properly.
USING
THE PARTICLE MEASURING SYSTEM:
USING
THE KEYBOARD:
LMB 08/03