BRIEF OPERATING INSTRUCTIONS FOR THE ZEISS EM 900
Please note: These
instructions are intended only for individuals who have had prior training on
this instrument. If at any point you have a question about the instrument's
operation or performance, please ask for assistance.
1. Turn on the water
valve. Turn on the power with the Mains Key Switch to 80 kV. Turn on the
nitrogen gas.
2. The pumping system
will now establish the proper vacuum in the column. When the vacuum is
established the "OPEN V3"
light will illuminate on the console and on the column next to the V3 valve.
Open valve V3.
3. Check the
following: Image mode on the microscope should be at "HR" and the magnification should be 3000. There should not be
a specimen in the beam path. If the scope was not left in this condition, make
the necessary adjustments.
4. The green beam
button should be illuminated. Push it
to turn on the high voltage. The red beam button should now be lit to indicate that the
high voltage is on. Now check to be sure the vacuum is still OK. Push the FIL button to turn on the filament. The
FIL button should now also be
illuminated.
5. Centering the
Illumination: Turn the small brightness
knob counter clockwise to crossover. Center the beam with the beam alignment knobs. Set the large brightness knob to position 5
(fifth dot). Now reduce the filament
heating slowly by turning the knob counter clockwise to undersaturate the
filament. Align the filament image if necessary with the filament precentering knobs.
6. Using the small brightness knob focus the
filament. Carefully (slowly) resaturate the filament by turning the filament heating knob clockwise. The
shadow should just disappear. Reset the large
brightness knob to position 3.
7. With the beam
still at crossover center the beam with the beam alignment knobs. Now use the small brightness knob to go back and forth through crossover. If
the beam spot does not expand concentrically, adjust the condenser aperture alignment as demonstrated in class. When the
alignment is complete, you should be able to turn the small brightness knob clockwise
and the beam should spread to evenly fill the screen. You will always work with
the beam between crossover and to the clockwise direction or
"low side"
of crossover.
8. Aligning and
Stigmating the Imaging System: Turn the large
brightness knob completely counterclockwise to position 1. Turn the small brightness knob clockwise to
overfocus. Set the imaging mode to "D".
With the small brightness knob and
the small
focusing knob adjust for the caustic figure. The P1 stigmator will help you get the caustic figure as close to a 3
pointed star as is possible. Return the large
brightness knob to position 3 and the image mode to "HR".
9. Centering the
Objective Aperture: Insert a specimen into the beam with the imaging mode still
at "HR". Be sure a piece
of tissue or plastic (not a grid bar or an empty space) is in the path of the
beam. Switch to imaging mode "D".
Focus the diffraction spot to its minimum size with the small fine focus knob and the small
brightness knob. Insert and center the objective aperture (the cluster of
small apertures is the third position in). Switch the imaging mode back to
"HR". Reduce the
brightness and adjust the magnification as needed.
10. ONLY IF NEEDED: Correcting
Objective Astigmatism: Image mode should be on "HR" and you will need
to use the holey grid. Adjust the microscope to about 25-50% higher
magnification than you plan to use. With a specimen in the beam, center a small
hole on the focusing screen. Focus the hole. Now overfocus the hole. You will
see the black Fresnel fringe. Check to see if the fringe is even all around the
hole by going back and forth through focus a few times. If the fringe is
uneven, use the objective stigmator
controls (carefully) to adjust it.
Special Notes:
IF YOU NEED TO LEAVE THE ROOM FOR ANY
REASON OR FOR ANY PERIOD OF TIME BE SURE THAT YOU DEFOCUS THE BEAM AND TURN OFF
THE HIGH VOLTAGE.
TO
CHANGE A SPECIMEN - Turn off the beam and close V3. Remove the specimen
holder from the microscope by turning the airlock counter clockwise. Remove the
holder using the appropriate tool from the kit. DO NOT TOUCH THE CARTRIDGE OR
THE SLIDE. Close the airlock slightly while you insert the grid into the
specimen holder. REMEMBER THAT TURNING THE CAP TO THE LEFT WILL LOOSEN IT - TO
THE RIGHT WILL TIGHTEN IT. MAKE SURE THE CAP IS NOT CROSS-THREADED, TOO TIGHT
OR TOO LOOSE. Reopen the airlock and insert the specimen holder with reference
to the locking pin. Close the airlock by turning it completely clockwise. PUSH
IN THE PUMP button on the console to
evacuate the airlock. Disengage the airlock rod and slide it into the three
o'clock position. Wait until the LOCK IN
light goes on and then about 30 seconds more. Then continue to insert the
specimen. You may leave the specimen in the airlock, but out of the beam as you
do your initial alignment. It will be ready to insert when you need it.
TO
TAKE A PHOTO load the film as demonstrated in class. Adjust the numbering
as necessary for your records. Focus your specimen using binoculars on the
focusing screen. The small screen is also the photometer, so leave it in place
as you adjust the brightness so that the exposure time will be 2 seconds. Bring
both screens up at the same time. Push the lit exposure button. The light will
go out. When it comes back on the screens should be lowered.
TO TURN OFF THE MICROSCOPE:
When you have finished microscopy for
the day:
1. Spread the beam by
turning the small brightness knob clockwise (so the beam is not at crossover).
Reduce the magnification to 3000.
2. Turn off the
filament by pushing the FIL button. Now turn off the high voltage by pushing
the RED beam button.
3. Remove your
specimen as demonstrated in class and replace the specimen cartridge in the
airlock as described previously. The cartridge may be left in the outermost
position as demonstrated in class.
4. Close valve V3.
5. Turn the
microscope key to OFF. Wait for about 3 minutes until you hear the dry nitrogen
gas fill the column, then turn off the nitrogen gas cylinder. Turn off the
water valve.
6. Remember to log
your filament time and the amount of film you used in the log book.
7. The room should be
NEAT and LOCKED when you are done.
LMB 3/02