GLOVEBOX USAGE |
Note: You cannot use the glovebox until you have been formally trained by Dr. Daley and he has signed-off on your use of the box. Overview: ("click here if you want to go straight to the general rules and glovebox operation directions) Even though the box is sealed and the atmosphere is circulated over a catalyst to keep it inert, there are still precautions that must be taken when working. Cross contamination of samples due to poor technique is problematic in particular when using volatile compounds and/or solvents. There are two common styles of use of gloveboxes:(i) one where the box is used solely to store, weigh, and transfer air-sensitive reagents without using any solvents or volatile liquids, and (ii) one where the former operations are performed as well as the use of solvents and performing of reactions that would be carried out using Schlenk techniques. Our glovebox is generally only used for the setup stage and product transfer stage of reactions (weighing compounds, transferring them to flasks or vials, dissolving them in solvent, etc.). However, on occaisson it can be used to perform entire synthetic operations including reactions in solvents, work-up, and preparation of samples for spectroscopy. Not all reagents and solvents are acceptable for use in the glovebox. The "box atmosphere" is usually continuously deoxygenated over a copper catalyst. Certain volatile chemicals such as halogenated compounds and strongly coordinating species such as phosphines and thiols can be problematic because they irreversibly poison the copper catalyst. Thus, in general we try not to use these in the glovebox and when we do it is for a minimal amount of time and the materials are immediately removed from the box as soon as possible. Please remember this as the copper catalyst is expensive and thus we do not want to have to frequently replace it! Always consult Dr. Daley when using the glovebox to ensure it is maintained in top condition. If at any time you notice an error on this page or if you have any questions, please contact Dr. Daley. Our glovebox along with labeling for ease of locating parts when reading the technique notes below.
GENERAL RULES AND PRACTICES
Read these before you use the glovebox to refresh your memory. See picture above for reference to the labels noted in the instructions below.1. Standard Settings and changing settings on the glovebox: To change glovebox settings, you use the control panel keypad located on the upper right of the box. The only settings you should be changing are (i) the light setting (on or off), (ii) the box pressure limits (settings for when using the box and when it is idle), (iii) the circulation settings (when using Difficult Chemicals and Solvent or an air-leak is found in the box), or (iv) turning on/off the oxygen and water sensors. Before you think about changing a setting, you need to know the Standard Settings, which are listed below along with what can/should be changed when using the box. Glovebox light: (i) when not in use - OFF; (ii) when in use - ON Glovebox pressure limits: (i) when not using the box - lower pressure set to 5 psi and upper pressure set to 10 psi; (ii) when using the box - lower pressure set to 3 psi and upper pressure remains set to 10 psi. If the glovebox pressure drops lower than atmospheric pressure too quickly, the box may not be able to compensate quickly enough, resulting in the possibility that ambient air could enter the box and thus raise the oxygen levels. Also note that if you are working and you hear the box constantly purging and filling it means you are likley (i) moving too quickly causing the pressure to raise or lower over the limit settings, or (ii) the settings on the box are too narrow (i.e. you do not have it set to 3 and 10 psi), or most likely (iii) you are not taking care to correlate the movements of your arms: when you reach in farther with one arm, you should be pulling the other one out slightly to compensate and keep the overall box pressure constant. Glovebox circulation: this is to remain ON at all times except the following circumstances. Glovebox oxygen and water sensors: thess are to remain ON at all times except the following circumstances. 2. Procedure in case of leak/hole found in box: 3. Enter and exit the box slowly: The box is not perfect, there are small holes that let ambient atmosphere
into the box. Therefore, the box should always be kept under a positive pressure. To change glovebox settings, you use the control panel keypad located on the upper right of the box. The settings to be used are: (i) when not using the box - lower pressure set to 5 psi and upper pressure set to 10 psi; (ii) when using the box - lower pressure set to 3 psi and upper pressure remains set to 10 psi. If the glovebox pressure drops lower than atmospheric pressure too quickly, the box may not be able to compensate quickly enough, resulting in the possibility that ambient air could enter the box and thus raise the oxygen levels. Also note that if you are working and you hear the box constantly purging and filling it means you are likley (i) moving too quickly causing the pressure to raise or lower over the limit settings, or (ii) the settings on the box are too narrow (i.e. you do not have it set to 3 and 10 psi), or most likely (iii) you are not taking care to correlate the movements of your arms: when you reach in farther with one arm, you should be pulling the other one out slightly to compensate and keep the overall box pressure constant. 4. Using the Antechamber (port): Reduce the amount of time the antechamber is left open by having all items you wish to bring in the box ready! Also, if someone else is using the antechamber, check with that person before using the antechamber. Keep it Clean: Return the glovebox better than when you found it.
LABEL EVERYTHING OR IT GETS THROWN AWAY! Store Vials Properly: Loose vials on the floor clutter an already limited work space. Please store all loose vials in the large, labeled crystallization dish. Vials that contain your samples must be stored on the shelves in one of the sample vial trays. If the trays are full, in consultation with Dr. Daley, remove unnecessary vials to make room or get a new tray. If we are running low on vials in the box, replace them with ones from the oven and store them in the label crystallization dish. Waste Handeling: All waste should be immediately removed from the box! If you use a chemical or solvent and leave the
waste lying around you are still contaminating the box atmosphere with whatever you used. We have a ziplock bag for used Kimwipes, etc. and a glass waste bottle for solvent waste. Each of these must be removed from the box after you have used it or if multiple people are using the box in one day, then the waste can be removed when the last person using the box is leaving for the day. Check the Balance: They are sensitive instruments. When they are not level, the weights they read are not accurate. The floor of the glovebox is uneven. When you move the balance, it is likely to not be level. If you want the weights you record to have useful physical meaning, then you don't want this. Before weighing make sure the balance is leveled by looking at the oil bubble leveler on the back of the balance (you will likely need a mirror for this). Keep the balance door closed at all times! This helps to prevent miscellaneous dirt from contaminating the balance which, of course, affects the weights you record.
Open Containers: Open solvent bottles, reactions (including venting with a needle stuck in the septum) are very bad in the glovebox because the atmosphere saturates in solvent vapors. This then leads to the catalyst bed becoming saturated in solvent vapors, which leads to a shorter catalyst lifetime (remember the catalyst costs a lot!). Also, solvent contamination will be a problem (e.g. diethyl ether in the NMR specturm of your reaction that never used diethyl ether ... guess where it came from!). Therefore, the box must be purged after usage of solvents or volatile chemicals (not including amines, phosphines, thiols, etc. "Dealing With Troublesome Chemicals" below for details on these). Needles are Sharp: They are the single largest hazard to the gloves. If you want to use a septum with a needle through it to dry something under vacuum, please don't push the needle all the way into the septum as the needle will stick out past the end of the septum, just begging for a glove puncture. Vacuum and Solvent Removal from Reaction Flasks:
SETUP. Before you begin, make sure the glovebox green valve and glass stopcock valve inside the glovebox are closed! Then setup the trap under the glovebox (T1) is in the Dewer, clamped securely to the ring-stand, and hooked up to the external vacuum pump and glovebox hoses. See the figure below for visual aids. Please note the trap is NOT in the Dewer only so that you can see the hookup. The first thing to do is put the trap into the Dewer, do not add liquid nitrogen at this point, and then hooking up the hoses as shown.1. Turn on the external vacuum pump to evacuate the system. 2. Add liquid nitrogen to the Dewer trap. Fill it all the way up and refill it to the top as necessary when solvent is being pumoped off. It should never go lower than 2 inches from the top. 3. Open the green valve at the inside back-right of the glovebox to open the hose in the glovebox to the vacuum/trap. 4. Place the hose on the side-arm flask to be evacuated and as you would on the Schlenk line, and with the side-arm flask stopcock closed to the hose, open the glass stopcock of the vacuum line from the glovebox hose to evacuate the line up to the flask. NEVER HAVE THE GLASS STOPCOCK OF THE GLOVEBOX HOSE OPEN WITHOUT BEING CONNECTED TO THE ITEM THAT IS TO BE EVACUATED! 5. Now you are ready to go! Slowly open the stopcock on the side-arm, while stirring the solution, to evaporate the solvent. 6. Once you have removed all the solvent to the point you desire, close the stopcock valve to the side-arm AND the glass stopcock of the glovebox connection. Note: DO NOT LEAVE THE ROOM WHILE THE EXTERNAL VACUUM PUMP IS ON! 7. Remove the hose to the side-arm and then slowly open the side-arm stopcock to re-pressurize the flask. SHUTDOWN. Once you are done using the vacuum (or the trap is full), close both valves inside the glovebox to the vacuum hose (green and glass stopcock valves). Disconnect the hose on the trap to the vacuum pump and then turn off the pump. Disconnect the hose from the trap to the glovebox line and then remove the trap and empty the contents to the waste bottle in the our Main Lab. Note: DO NOT LEAVE THE PUMP ON IF: you are not using it or the trap is not filled with liquid nitrogen when you are using it (the trap contents will all end up in the pump oil). VIDEO HERE and/or Pictures Dealing With Troublesome Chemicals (including volatiles): Amines, phosphines, and thiol compounds are very problematic. Not only do they bind very well to the catalyst bed, but
they can leave a white film on everything in the box. When using these compounds (or any other volatile chemical including solvents) follow these steps: 1. Shut off the blower and close off the catalyst bed. This is done by hitting the "Circulation" button on the panel attached to the upper right side of the glovebox (P1) so that the green light is off. When you switch it off you will immediately hear a winding down of the fan from the blower. If you do not hear it, stop using the box and contact Dr. Daley immediately as there could be a serious problem. 2. Use the chemicals as required taking care to be as quick as one can be while being safe and clean. 3. After using these troublesome compounds, reseal the bottles/vials tightly with black tape and/or parafilm. 4. Purge the box. Change the pressure settings of the box to a high of 14 psi and a low of 10 psi, changing the high setting first. Use the wrench to open the screw valve port on the left front side of the glovebox and remove the screw. Place the purge septum adaptor over the open screw port. Remove the protective septum from the inside openning to the screw port. At this point the box will begin purging. Let the box purge for a minimum of 15 minutes (ideally go for about 30 minutes). Once the purging is completed, place the protective septum over the inner screw port opening, replace the purge septum adaptor with the screw, and tighten the screw snuggly - do not over-tighten it as this will cause damage. If you are unsure, please ask Dr. Daley for help. 5. Turn the blower back on by pressing the "Circulation" button until the green light is on. Again, you should hear the blower fan start up. If you do not, please contact Dr. Daley immediately. 6. Material that came into contact with these troublesome compounds must be removed from the box immediately. Everyone is responsible for their own trash. Bringing Flasks Into the Glovebox:
Flasks and other containers being brought into the glove box are subject to high vacuum, and therefore need to be appropriately sealed to prevent them from busting open in the antechamber. (i) When bringing an empty flask into the glove box: take it directly from the oven to the glovebox (put a stirbar that has also been in the oven overnight if needed) antechamber along with a sepctum. Do not seal the flask with the septum, just place the septum next to the flask in the antechamber and then use the purge-fill cycles to bring the equipment in. (ii) When bringing a sealed flask that contains no solvent into the glove box: simply ensure that it has been completely evacuated on the Schlenk line by keeping it under vaccum for at least 1 minute and then sealing the flask. Examples of these flasks would be reaction flasks that have had solids in them that were (i) starting materials massed-out in the lab and then placed in then cycled on the Schlenk line or (ii) reaction products/residues that remain after filtration or evaporation of the solvent on the Schlenk line. (iii) When bringing a sealed flask containing solvent into the glove box (e.g. Diethyl ether): you must be very careful here! You can only do this after you have been formally trained by Dr. Daley. The step to perfrom are: Maintenance:
1. Change the pump oil: This needs to be done every two months when the box is used regularly. At minimum it should be changed every six months. Dr. Daley will take care of this.2. Patching Holes: If you poke or notice a hole in the glove box (typically the in a glove) stop using the glovebox and contact Dr. Daley immediately. Holes on the gloves can be fixed with a "Bike Tire Repair Kit" in many cases but only Dr. Daley can do this. If the hole is in a glove and you cannot find Dr. Daley (or Dr. Clark or Dr. Citeau) then please use the protective glove port screw cap to seal the port of the glove with the hole. The cap is located on the top shelf - you pull it into the glove port opening with your hand on the knob. Once in place you turn the knob until the port is sealed. The glove will immediately deflate but do not worry as the rest of the box is protected. Hoever, do not use the box until the glove has been fixed.
|